ASTM F86 Cleaning/Passivation

X-medics’ tantalum beads are cleaned/passivated according to ASTM F86 (Standard Practice for Surface Preparation and Marking of Metallic Surgical Implants) by an ISO 13485 certified company located in Switzerland.

This treatment ensures that the tantalum beads are free from foreign metal particles, grease and other possible contaminants arising during manufacturing and processing.

The validated ASTM F86 cleaning procedure of the tantalum beads occurs in several steps:

  1. Pre-cleaning in hot alkaline solution
  2. Cold rinse in tap water
  3. Cold rinse in osmosis water
  4. Nitric acid passivation
  5. Acidic rinse
  6. Cold rinse in osmosis water
  7. Hot rinse in osmosis water
  8. Hot rinse in demineralised water
  9. Final hot rinse in demineralised water
  10. Drying in hot air stream

The alkaline cleaning removes grease, oil, dirt and other loose particles from the surface of the tantalum spheres.

Passivation in nitric acid removes embedded particles from production – typically iron filings from tooling (steel). The nitric acid dissolves iron particles selectively, as tantalum material is non-reactive with nitric acid.

Ultrasonic baths are used during the treatment to ensure efficient removal of particles from the surface of the tantalum balls.

Read more about ASTM F86: Standard Practice for Surface Preparation and Marking of Metallic Surgical Implants